A detailed description of the setup and of the procedure used for the large-scale quality test of the 154 560 limited streamer tubes which compose the CHARM II calorimeter is given. The size of the project has imposed the adoption of a procedure that is simple and fast, yet dependable. The high reliability of the test has been proven by the stability of the system after 20 months of continuous operation

Large-series Test of Limited Streamer Tubes

GORINI, Edoardo;
1987-01-01

Abstract

A detailed description of the setup and of the procedure used for the large-scale quality test of the 154 560 limited streamer tubes which compose the CHARM II calorimeter is given. The size of the project has imposed the adoption of a procedure that is simple and fast, yet dependable. The high reliability of the test has been proven by the stability of the system after 20 months of continuous operation
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.